ALTANA Haze-Gard Plus Short Instructions User Manual
English, Haze-gard plus operation, How to calibrate
269 020 584 E 0806
Additives & Instruments
A member of
Measure what you see.
BYK-Gardner GmbH · Lausitzer Strasse 8 · 82538 Geretsried · Germany · Tel. +49-8171-3493-0 · Fax +49-8171-3493-140 · www.byk.com/instruments
Please note the detailed operating instructions in English on CD.
ENGLISH
haze-gard plus
Operation
On the left Side:
•
Main Switch
on/off
•
Voltage Selector
•
Fuse
Connection for:
•
Power Cable
•
Foot Switch
•
PC
How to Calibrate?
Calibration is necessary at the first
startup and at a regular interval
(two months recommended):
•
Press set up button to call up
the calibration routine
•
Cover haze-port with dark
standard and press operate
•
Remove cover from haze-port
and start the 100%
measuremen
•
Place the clarity standard at the
clarity port and press operate
operate button:
Measure
set up button:
SET-UP MENU
reference button:
Measure reference in Continuous
mode
Cursor keys:
Used in SET-UP mode to select
individual menu items and set
values.
In any measuring mode, these keys
are used to switch statistics
functions on/off, and to delete
values from statistics memory
SET UP MENUE
¢
Calibrate
£
Print Statistic
£
Print Online:
off
£
Cont. Mode:
off
£
Language
USA
Escape:
SET UP
delete all
delete last
statistic
result
on/off
on/off
operate
set up reference
trans-
mittance
haze
clarity
delete all
delete last
statistic
result
on/off
on/off
H
1 4 . 7
clarity-port
haze-port
Calibration
Zero Calibration
Calibration
100% Calibration
Calibration
Place clarity stand
75.5
at clarity-port