11 advanced configuration – ALTANA Wave-Scan II User Manual

Page 37

background image

36

English

11

Advanced Configuration

Correction

Defects on the sample surface, such
as scratches or craters, can cause
major errors in measurement values.
When Correction is activated, the
affected scan areas are cut out and
the measurement values are calculated
from the sanitized data. Pressing
“operate” turns Correction on and off.

Plausibility Control

An option for comparing the corrected
and uncorrected measurement value.
The greater the difference between the
corrected and uncorrected data, the
more critical is the surface defect. If
the difference is greater than 20 %, the
measurement will be evaluated as a
faulty measurement. An error message
appears and a new measurement is
then needed.
Pressing “operate” turns Plausibility
Control on and off.