5. Press the flashing
Start Test
button and proceed immediately to step 6.
6. Lift the locking pin and release the weight by lowering your hand
quickly and smoothly.
The CMRV-4500 will
measure the speed of rotor
rotation by timing the
revolution(s) of the pulley-
wheel as the weight falls.
NOTE
The total number of rotations
measured by the CMRV-4500
during the viscosity test depends
on the viscosity of the sample
and the values chosen in the
Advanced Settings box. Default
values for the Advanced Settings
correspond to the specifications
of the ASTM method selected for
the profile.
7. When the Viscosity LED on
the CMRV front panel stops
flashing, indicating the
conclusion of the viscosity
test, remove the hook-weight
from the string.
CAUTION
Do not remove the weight while the Viscosity LED is still flashing.
NOTE
To view test results as soon as they have been recorded in the data-
base, open an MRV Data Table analysis with the Use Dynamic Update
option (see Chapter 13).
8. To test additional cells, repeat steps 1-7 above, testing in order from
left to right.
After all viscosity tests have been concluded, select
End Profile
from the
Instrument View window to terminate temperature control of the CMRV.
NOTE
No further testing can be conducted after temperature control has been
aborted.
Notes on viscosity testing
Do not let the large hook-weight swing back and forth during testing or
the apparent viscosity measurement may be distorted.
Viscosity data
The computer determines the apparent viscosity, along with the time
required for rotation of the rotor. Time and temperature data for the test is
recorded in the associated .ttd file.
Placing the large hook-weight in
preparation for the test.