Temperature:

Click

to select the desired temperature for calibration.

Mass:

Click

to select the desired test mass for calibration.

Known Viscosity:

Enter the known viscosity of the standard, as printed

on the standard bottle label.

Calibrate ... :

After you have entered the calibration data described

above, click the Calibrate ... button to access the CMRV

Cell Calibration

window.

The Calibrate Cell Constants window

Reset Constants to Original:

Click the

Reset Constants to

Original

button to restore constants to the last saved values.

Abort Temperature Control:

Click the

Abort Temperature Control

button to abort CMRV temperature control. The calibration procedure
will be halted.

Abort Calibration:

Click the

Abort Calibration

button to abort CMRV

calibration. The individual cell calibration will be halted if the

Start Test

button on the CMRV instrument has not yet been pressed.

Save Data:

Click the

Save Data

button to save the current cell con-

stants to the current instrument configuration.