Cannon Instrument CMRV-4500 User Manual
Page 59

53
CANNON
®
Mini-Rotary Viscometer CMRV-4500 Instruction & Operation Manual
Version 1.1e—October, 2011 •
CANNON
®
Instrument Company
2139 High Tech Road • State College, PA • 16803 • USA
Temperature:
Click
to select the desired temperature for calibration.
Mass:
Click
to select the desired test mass for calibration.
Known Viscosity:
Enter the known viscosity of the standard, as printed
on the standard bottle label.
Calibrate ... :
After you have entered the calibration data described
above, click the Calibrate ... button to access the CMRV
Cell Calibration
window.
The Calibrate Cell Constants window
Reset Constants to Original:
Click the
Reset Constants to
Original
button to restore constants to the last saved values.
Abort Temperature Control:
Click the
Abort Temperature Control
button to abort CMRV temperature control. The calibration procedure
will be halted.
Abort Calibration:
Click the
Abort Calibration
button to abort CMRV
calibration. The individual cell calibration will be halted if the
Start Test
button on the CMRV instrument has not yet been pressed.
Save Data:
Click the
Save Data
button to save the current cell con-
stants to the current instrument configuration.