CAUTION
Make sure that the
assembly is seated
securely on the
track.
3. Replace the
left front
screw. This
will prevent
the assembly from sliding off the track.
4. Plug the free end of the pulley-wheel sensor wire into the jack
labelled
WHEEL
on the front of the CMRV-4500 controller.
NOTE
To disconnect the sensor, pull it out by the knurled portion of the plug.
Installation notes
Download button
The CMRV-4500 is capable of receiving firmware updates from the
computer via the
DOWNLOAD
button on the rear of the electronic
chassis. Should such an update be required,
CANNON
®
will provide
detailed instructions regarding the download procedure for updating the
firmware.
NOTE
If the DOWNLOAD button is pressed inadvertently, switch off the CMRV-
4500 power for at least four seconds, then
restore power to the unit.
Setting the CMRV instrument address
When installing/connecting a new CMRV
instrument, you must set the instrument
address using the
SELECT ADDRESS
dial on the rear of the CMRV controller. This dial offers 16 settings (0-9,
A-F).
Procedure
To set the address, rotate the dial (see photo) to a setting not currently in
use by other CMRV instruments.
CAUTION
The MRVW software will not function correctly unless each networked
CMRV instrument is set to a different address (see APPENDIX D for
multi-unit configuration).
Pulley-wheel assembly mounted on slide track
Address selection knob