Logging in, Checking instrument settings – Cannon Instrument CMRV-4500 User Manual
Page 29

23
CANNON
®
Mini-Rotary Viscometer CMRV-4500 Instruction & Operation Manual
Version 1.1e—October, 2011 •
CANNON
®
Instrument Company
2139 High Tech Road • State College, PA • 16803 • USA
Checking Configuration data
Follow the procedures in the next several sections of this chapter to select
and verify the instrument and calibration settings to ensure that they
conform to the actual characteristics of your
CANNON
®
instrument.
Configuration protection
To check the configuration settings for your instrument(s), you must log
in to the security system as a manager. The software is installed with a
default Manager account. This account has no password, allowing any
operator access to manager-level software functions as long as the
password is not activated/changed. If you would like to engage the full-
release security options, see Security Options, this chapter, for instruc-
tions.
Logging in
1. Use your mouse to click
Main
from the VISCPRO
®
menu bar.
2. Click
Log In
from the
Main
menu options.
3. Click on the
(arrow) on the right side of the
User Name:
list box
to display the list of registered users.
4. Click
Manager
. Do NOT enter a password!
5. Click
OK
. The
Log In
window will close automatically and you will
be logged in as management personnel.
Checking Instrument Settings
1. Use your mouse to click (select)
Configure
from the VISCPRO
®
menu bar.
2. Select your instrument from the list of available instruments (there
may be only one instrument in the list).
3. Select
Instrument Settings
from the list of configuration options.
The
Instrument Settings
window will appear.
You will use the
Instrument Settings
window (see graphic following) to
describe and control MRV instrument operational features. These settings
affect the instrument as a whole. Check the instrument settings for your
instrument per the instructions, and make any necessary changes: