beautypg.com

Running profiles – Cannon Instrument CMRV-4500 User Manual

Page 65

background image

59

CANNON

®

Mini-Rotary Viscometer CMRV-4500 Instruction & Operation Manual

Version 1.1e—October, 2011 •

CANNON

®

Instrument Company

2139 High Tech Road • State College, PA • 16803 • USA

CHAPTER

11

RUNNING PROFILES

NOTE

The guidelines in this section are not intended to supercede any test
method. Always refer to the method for validation of any test procedure.

Test preparation

To run a temperature profile, first prepare the CMRV-4500 for testing
(see Chapter 6). Make sure that the Air/Water Heat Exchanger

POWER

switch is on and that the unit is operating normally.

Procedure

1. Place 10 ml of test oil in each of the viscometric cells to be used. Any

number of cells, one to nine, may be used but all test oils for the run
must be for the same test temperature. Use a syringe or small plastic
cup to fill the cells.

NOTE

If testing with fewer than nine cells, fill unused cells with 10 ml of a clean
oil prior to running the test.

2. Load the VISCPRO

®

software and click

Main/View Instrument

.

Select the desired instrument type (MRV) and configuration from the

View Instrument

window. Then click

OK

.

3. Record the sample ID for the first calibrated MRV cell by clicking in

the Sample ID field and typing the sample name. Then press the

Tab

key to move the cursor to the next Sample ID field (or just click in
the desired ID field). Repeat the process until ID information has
been entered for all test samples.

4. (optional) Save your MRV instrument configuration by clicking

Main/Save Instrument

. Select your instrument type (MRV) and

enter a name for the configuration in the

Save As:

field. Then click

OK

to save the configuration.

NOTE

You may change the Sample ID at any time prior to initiating the Yield
Stress or Viscosity tests. If you change the sample information, you may
wish to save the configuration (see step 4, above) before closing the
Instrument View window or your changes may be lost.

5. Carefully insert the rotors into their corresponding test cells, making

sure that the rotor tip at the bottom of the shaft slides into the recess
at the bottom of the cup.

CAUTION

Do not drop the rotors into place or you may damage the rotor tips.

6. Secure each rotor in position with a rotor bearing pin.