Cannon Instrument CMRV-4500 User Manual
Page 24

18
CANNON
®
Mini-Rotary Viscometer CMRV-4500 Instruction & Operation Manual
Version 1.1e—October, 2011 •
CANNON
®
Instrument Company
2139 High Tech Road • State College, PA • 16803 • USA
4. Replace the external housing on the Air/Water Heat Exchanger and
secure it with the six screws previously removed. If necessary, reseat
the gasket around the reservoir opening with a small screwdriver.
5. Add water-antifreeze mixture to the Exchanger per the earlier
instructions in this chapter to complete the water-antifreeze flush and
drain procedure.
FHP MOTOR
NOTES
For rountine maintenance and improved bearing life expectancy, a few
drops of non detergent “twenty weight” oil can be added every 12 months.
Fans should be cleared of dust/other material on a regular basis.