beautypg.com

Starting a profile – Cannon Instrument CMRV-4500 User Manual

Page 67

background image

61

CANNON

®

Mini-Rotary Viscometer CMRV-4500 Instruction & Operation Manual

Version 1.1e—October, 2011 •

CANNON

®

Instrument Company

2139 High Tech Road • State College, PA • 16803 • USA

cover (Step 9.2.6 in the ASTM D 4684-98 method and step 9.1.6 in
the ASTM D 3829 method). The tubing may be removed, along with
the cover, prior to viscosity/yield stress testing.

14. Whatever the methodology, the purge should continue throughout the

temperature profile using dry gas at a flow rate between 20 and 30
milliliters per minute.

Starting a profile

1. If the

View Instrument

window is not open, click

Main/View

Instrument

and select the desired instrument type (MRV) and

configuration from the

View Instrument

window. Then click

OK

.

Polling instruments

If you are unable to select your CMRV from the list of available instru-
ments in the
View Instrument window, make sure the instrument is
turned on and warmed up (see Chapter 6); then select Poll for Instru-
ments from the Main menu.

2. From the MRV

Instrument View

window, click the

Upload Profile

button to open the

Upload Profile

window.

3. Select a

Standard

or

User-Defined

profile from the available

profiles in the drop-down box, then click

OK

.

NOTES

To run samples using -20°C or -25°C calibration data per ASTM specifi-
cations, ensure that the correct profile is selected from the ASTM Stan-
dard Profile laboratory. Standard Profiles with only one temperature
indicated are run using calibration data obtained at -20°C. For profiles
with two temperatures indicated, the first temperature refers to the final
test temperature for the profile and the second refers to the desired
calibration temperature for that profile.

(-10°C profile using -20°C calibration data)

You will not be able to select and run ASTM profiles in cells for which
calibration data is unavailable at the specified temperature. The ASTM
profiles will not appear in the drop-down list until the number of calibra-
tions required by the method are performed.

4. Click the

Run Profile

button to open the

Run Profile

window. The

on-board CMRV microprocessor will control the temperature until
the profile has been completed. A computer connection is no longer
required until it is time to initiate viscosity/yield stress testing.

5. Select the desired

End Time

for the profile (this is the time you

would like to start testing) and click

OK

. The Start Test button will

begin blinking to indicate that profile timing has begun. The profile
will be initiated at the appropriate time.

NOTE

The default End Time is the time that the temperature profile would be
completed if the profile is started immediately.