27

CANNON

®

Mini-Rotary Viscometer CMRV-4500 Instruction & Operation Manual

Version 1.1e—October, 2011 •

CANNON

®

Instrument Company

2139 High Tech Road • State College, PA • 16803 • USA

Procedure

1. Click

Log In

from

Main

.

2. Select your Manager or Service level user name from the

User

Name

list box. Input the correct password in the appropriate field.

3. Click

OK

.

4. Click

Update User Information

from

Main

.

5. Select the desired user from the

User Name

list box.

NOTE

To delete a user, just click Remove User after selecting the user name.
The account will be immediately and permanently removed.

6. Enter and verify the desired password, and select user security status

using the appropriate list boxes.

7. Retype the password in the

Confirm Password

field.

8. Click

Add or Update User

to save your changes.

9. Click

Done

to exit the

Change User List

window.

To use the

Log Out

feature, click

Log Out

from

Main

. The current user

will be logged out of the security list of authorized technicians and
managers. The software will automatically reset to the lowest security
level. Any ongoing test operations will continue.