The Instrument Settings window

4. Use the ID field to input instrument identification information using

up to 30 alphanumeric characters.

The remaining fields in the

Instrument Settings

window are non-

editable (information is obtained via serial communication with the
CMRV instrument). The

Model:

field will indicate the model of your

instrument. The

S/N:

field indicates the serial number from the label

on the rear service panel. The

Firmware:

field indicates the version

for the current instrument firmware. The

Port:

field indicates the

current communications port for the RS-232 cable connecting your
computer to the CMRV-4500 instrument.

5. When you have entered all settings, click

OK

.

Checking initial calibration

After you have completed the CMRV instrument calibration for tempera-
ture and cell constants (see following chapters), follow the procedure
below to verify the initial calibration settings.

1. From the VISCPRO primary window, click

Configure/MRV/View

Cell Constants

. The

View Cell Constants

window will open.

2. Ensure that Calibration data, including constants and viscosity, is

available for each cell. Then click

OK

to close the

View Cell Con-

stants

window.