The Instrument Settings window
4. Use the ID field to input instrument identification information using
up to 30 alphanumeric characters.
The remaining fields in the
Instrument Settings
window are non-
editable (information is obtained via serial communication with the
CMRV instrument). The
Model:
field will indicate the model of your
instrument. The
S/N:
field indicates the serial number from the label
on the rear service panel. The
Firmware:
field indicates the version
for the current instrument firmware. The
Port:
field indicates the
current communications port for the RS-232 cable connecting your
computer to the CMRV-4500 instrument.
5. When you have entered all settings, click
OK
.
Checking initial calibration
After you have completed the CMRV instrument calibration for tempera-
ture and cell constants (see following chapters), follow the procedure
below to verify the initial calibration settings.
1. From the VISCPRO primary window, click
Configure/MRV/View
Cell Constants
. The
View Cell Constants
window will open.
2. Ensure that Calibration data, including constants and viscosity, is
available for each cell. Then click
OK
to close the
View Cell Con-
stants
window.