Instrument Settings window button options:
saves the current instrument settings and exits the
Instrument Settings
window.
closes the
Instrument Settings
window without saving
any configuration changes.
Profile Designer
The
Profile Designer
option opens the
Profile Designer
window,
enabling selection and editing of temperature profiles. For information
using the Profile Designer and Profile Editor, see Chapter 14, Designing
Customized Profiles.
View Cell Constants
The
View Cell Constants
option opens the
View Cell Constants
window. You can view current cell constants from this window, and you
can edit them if you are logged in to VISCPRO
®
as a Manager.
The View Cell Constants window
Viewing Cell Constants
To view cell constants for a specific temperature, click
from the
temperature field to display the valid calibration temperatures. Then click
the desired temperature to select it. Do the same for the test Mass.
Editing Cell Constants
To edit Cell Constants, follow the instructions below:
1. Log into VISCPRO
®
as a Manager.
2. Open the View Cell Constants window.
3. Select the desired temperature and test mass (see Viewing Cell
Constants, above).