Mini-Rotary Viscometer CMRV-4500 Instruction & Operation Manual

Version 1.1e—October, 2011 •

CANNON

®

Instrument Company

2139 High Tech Road • State College, PA • 16803 • USA

Restore procedure

To restore instrument settings from a saved configuration:

1. Click on

View Instrument

from the

Main

menu options.

2. Select the desired instrument from the left list box.

3. Select the desired saved configuration from the right list box.

4. Use the check boxes to select the desired configuration elements to

be restored.

Restore Instrument Settings

restores instrument properties from

the saved configuration.

Restore Tray Settings

is not applicable to the CMRV instrument.

Restore Sample Information

restores all sample ID setup informa-

tion for the Instrument View window from the saved configuration.

5. Click

OK

. If you clicked the check box to restore instrument, tray or

sample settings, the current instrument will be updated with the
selected settings.

NOTE

If you have updated the settings, you may wish to save the current
instrument/tray/sample configuration by selecting
Save Instrument from
the
Main menu, inputting the configuration name in the Save As: text
box, and clicking
OK.

Calibrate Cell Constants

The Calibrate Cell Constants

option from the

Service

menu initiates a

software-guided cell calibration procedure. For detailed information on
calibrating the CMRV cells, see Chapter 8.

The Calibration Parameters window
The

Calibration Parameters

window options are described briefly

below: