16. When acceptable values have been obtained, click
Abort Tempera-
ture Control
. Then click
Save Data
to save the new calibration
constants. Do not omit this step or new calibration data will be lost.
17. Close the
Calibrate Cell Constants
window and then click
Main/
Save Instrument
. The
Save Instrument
window will appear.
18. In the
Save As
box, input a name for the new instrument configura-
tion (which includes the newly-calculated calibration constants).
Then click
OK
to save the configuration.
Printing calibration data
CANNON Instrument Company recommends generating a hard copy
record of calibration data after each calibration session. To print the
current calibration constants, click
Configure/Print Instrument Set-
tings
.