16. When acceptable values have been obtained, click

Abort Tempera-

ture Control

. Then click

Save Data

to save the new calibration

constants. Do not omit this step or new calibration data will be lost.

17. Close the

Calibrate Cell Constants

window and then click

Main/

Save Instrument

. The

Save Instrument

window will appear.

18. In the

Save As

box, input a name for the new instrument configura-

tion (which includes the newly-calculated calibration constants).
Then click

OK

to save the configuration.

Printing calibration data

CANNON Instrument Company recommends generating a hard copy
record of calibration data after each calibration session. To print the
current calibration constants, click

Configure/Print Instrument Set-

tings

.