beautypg.com

Brookfield DV3T Rheometer User Manual

Page 76

background image

Brookfield Engineering Labs., Inc.

Page 76

Manual No. M13-2100

V.4.5 Default Path

The Path is the location where data or tests are stored. The DV3T will allow for storage to the

Internal Memory or to an attached USB Flash drive. The path will also include any file structure

that has been created.

The Default Path identifies what location will be used as the initial location when Results are saved.

The Default Path defined in Reports - Configure Viscosity Test (saved test or unsaved test) will

take priority over the Default Path.

The DV3T can create file structure based on User ID and Test Name. The options provided for the

Default Path specify what format the data path should utilize. For example: set the Default Path

to Operator -> Test Name. The User is logged in as DAVID and the test to run is INK 5. The data

path will be defaulted as shown in Figure V-12.

Figure V-12

TIP: The Default Path can be overwritten at the time that data is saved.

V.4.6 Settings Reset

Settings Reset will return the DV3T settings (Device, User and Global) to the factory default.

Internal memory will not be affected. User accounts will not be affected.

V.4.7 Device Reset

DATA LOSS POSSIBLE!

Device Reset will return the DV3T to the factory default condition. All data, tests and Audit Trails

will be deleted from internal memory. All User accounts will be deleted.

TIP: Move data files and test files to a USB Flash Drive prior to selecting Device Reset.