Brookfield DV3T Rheometer User Manual

Page 62

background image

Brookfield Engineering Labs., Inc.

Page 62

Manual No. M13-2100

Communication is established when the Test Connection command button is pressed. You must

press this button every time the DV3T Rheometer is powered on. Be sure that the temperature

controller is turned on prior to pressing the Test Connection button.

The Default Temperature screen also allows the user to set a Default Temperature. The Default

Temperature is the set point value that will be used by the DV3T when a test (Viscosity Test or

Yield Test) is not in use. To set the Default Temperature first, enter a temperature value and second

check the Default Between Tests check box.

TIP: Use a Default Temperature when the test program includes a change in temperature and

set the value equal to the first temperature control point in the test. At the conclusion

of the test, the control device will be reset to the first temperature set point which will

minimize the time required between tests.

The Temperature screen also allows you to choose which temperature probe will be used for data

collection and temperature control. When the DV3T is controlling temperature through an external

device, there may be two temperature probes available for data collection: 1) the temperature probe

connected to the DV3T (Probe) and 2) the temperature probe associated with the temperature

control device (Controller). Press either Probe or Controller in the Temperature menu, figure the

first one above.

Printer Setup: The DV3T is configured for use with the Dymo 450 printer. No other printers can

be utilized with the rheometer. Printer Setup allows the user to choose from three

different configurations for print out.

Label Small: A small label (1.13inch x 3.5inch) that allows for printing of limited

data for a single data point. The label stock is available from

Brookfield; part number GV-1048.

Label Large: A large label (2.31inch x 4.0inch) that allows for printing of a

complete set of parameters for a single data point. The label stock

is available from Brookfield; part number GV-1049.

Receipt: A continuous roll of paper that allows for printing of multiple data

points. The paper stock is available from Brookfield; part number

GV-1047.

The Printer Setup is selected by touching the desired configuration, which will

result in a blue circle to appear to the left of the item. The OK command key must

be selected to confirm the setting.

AutoZero: AutoZero is an operation performed by the DV3T Rheometer automatically during

the power up sequence. This operation sets the instrument transducer to a correct zero

value. The zero value should not shift over time, however, if the user determines that

the zero is not correct, you can force an AutoZero to be performed without powering

down the rheometer. When the AutoZero function is selected, the DV3T is immediately

sets to AutoZero mode and the user is presented with the AutoZero screen (Figure II-2).

Upon completion, the user will be returned to the Device Setup menu.

Oscillation Check: The Oscillation Check is a check of the performance of the lower bearing of the

DV3T Rheometer. This lower bearing is either Point and Jewel or Ball Bearing.

Correct performance of the lower bearing is essential for proper operation of the

rheometer. A failed lower bearing will normally cause viscosity measurements

to be lower than expected.

An Oscillation Check is performed as follows:

• Remove the spindle and level the DV3T Rheometer

• Gently push up on the rheometer coupling.