The Command Keys include Clear, Save and Run.

Clear: Clear all data that has been entered into the test parameters

and restore the values to the factory default.

Save: Save the current test.

Run:

Run the current test.

The Test Parameter area includes many elements of the viscosity test as well as live measurements

of Torque % and Temperature. Temperature data will only be displayed if a Brookfield temperature

probe is connected to the DV3T Rheometer.

Torque:

A live signal from the rheometer.

Spindle:

The currently selected spindle. All viscosity, shear

rate, and shear stress calculations will be made based

on this spindle. The spindle number may be changed

by pressing the blue arrow.

Speed:

The currently selected speed of rotation. The

rheometer will operate at this speed once the RUN

command key is pressed. The speed may be changed

by pressing the blue arrow.

Temperature:

A live signal from the rheometer when a temperature

probe is attached (Brookfield part number DVP-94Y

or SC4-61Y).

End Condition:

Specify the condition that will end the test.

Data Collection: Specify the amount of data to be collected during

the test.

Instructions:

Create a message that the user will see when the test

begins.

Reports:

Define how the data will be viewed when the test is

complete.

QC Limits:

Define the limits for acceptable measurement data.

Density:

Define the density of the test sample. This information

will be used when kinematic viscosity units are

selected for display (see Section V.4.2)

II.5.2

Configure Yield Test

Yield measurements are made on the DV3T Rheometer through the Configure Yield Test function.

The user can select Configure Yield Test from the Home Menu

.