Brookfield DV3T Rheometer User Manual

Page 70

background image

Brookfield Engineering Labs., Inc.

Page 70

Manual No. M13-2100

Code

The code is the numerical value used to select the spindle. This value

must be unique and can not be the same as a standard Brookfield

spindle. The acceptable range for Code is 100-199.

Name

The Name is a unique value to describe the spindle. The Name will

be displayed in the Spindle field. The Name can be comprised of

alpha and numeric characters.

SMC

The SMC (Spindle Multiplier Constant) is used to convert the measured

torque to viscosity. The acceptable range of SMC is 0.001 – 500.
The SMC value can be determined through the use of the Range

equation found in appendix D. Some experimentation with the spindle

and a calibrated viscosity standard is required.
Range (cP) = TK * SMC * 10,000/RPM
The Torque Constant (TK) can be found in Table D-2 in Appendix D.

SRC

The SRC (Shear Rate Constant) is used to convert the set RPM to

shear rate and calculate Shear Stress. The acceptable range of SRC

is 0.001 – 500.
The SRC value can be determined if the shear rate characteristics of

the spindle geometry are known. Consult the Brookfield publication,

“More Solutions to Sticky Problems”, for a discussion about shear

rate and some recommended equations.
Shear Rate (1/s) = SRC * RPM

YMC

The YMC (Yield Multiplier Constant) is used to convert the measured

torque to yield stress.

Speed List: The selection of Speed in Configure Viscosity Test can be done by using either a

number pad or a scroll list. (This choice is selected through the navigation button in

the Set Spindle screen.) When scroll list is selected, the speed entry is restricted to

the classic Brookfield speed sets for the LV and RV/HA/HB rheometers. The contents

of the scroll list are defined in the Speed List menu. Each speed can be individually

selected or deselected through the use of a check box. Select or deselect all speeds

by using the Select All check box. Select or deselect all LV (or RV/HA/HB) speeds

by using the Select All LV check box. The classic Brookfield speed sets are defined

below.

LV:

0.3, 0.6, 1.5, 3, 6, 12, 30, 60

RV/HA/HB:

0.5, 1.0, 2.0, 2.5, 4, 5, 10, 20, 50, 100

Additionally, a new speed can be added to the list by selecting the New Speed command

key at the bottom of the display. The New Speed must be selected within the range

of available speeds on the DV3T: 0.0 – 250 RPM.
Speed entry through the number pad is never restricted.

TIP: To ensure that the correct speed is always entered, create a test that is saved

to the DV3T memory.

TIP: The selection of speed can be restricted by User Log In. Use this restriction

combined with a saved test configuration to ensure that the correct speed

is always used for critical tests. See Section V.4.2 for details of Users and

Access.