II.7 Out of Range

The DV3T Rheometer will give on screen indications when the measurement is out of range of

the instrument. When the %Torque reading exceeds 100% (over range), the display of %Torque,

Viscosity, and Shear Stress will be EEEE (see Figure II-7). If the %Torque value is between 0 -

9.9%, the data field label will flash. When the %Torque is below zero (negative values), the display

of Viscosity and Shear Stress will be ----.

TIP: Brookfield recommends collecting data only when the %Torque reading is between

10 - 100%.

Figure II-7

Measurement data should not be collected when the %Torque reading is out of range. The out of

range condition can be resolved by either changing the speed (reduce speed when reading is out of

range: high) or changing the spindle (increase the spindle size when the reading is out of range: low).

TIP: When comparing data, the test method is critical. Be sure that you know the proper spindle

and speed required for the test method. If readings are out of range, this condition should

be reported as the test result.