Brookfield DV3T Rheometer User Manual

Page 13

background image

Brookfield Engineering Labs., Inc.

Page 13

Manual No. M13-2100

II. GETTING STARTED

II.1 Power Up

The DV3T will go through a Power Up sequence when the power is turned on. The Rheometer

will issue a beep, present a blue screen, and finally show the DV3T About screen for 5 seconds.

The About screen is shown below and includes several critical parameters about the rheometer

including; rheometer torque (LV, RV, HA, HB, or other), firmware version number, model number

(DV3TLV for example) and the serial number.

Figure II-1

The About screen can also be accessed through the Settings Menu (see Section V.1).
The DV3T Rheometer will automatically transition from the About screen to the AutoZero screen.

TIP: When contacting Brookfield or your authorized Brookfield dealer for technical support or

repair services, please record the information on the About screen and include this detail

in your email.

II.2 AutoZero

The DV3T Rheometer must perform an AutoZero prior to making viscosity measurements. This

process sets the zero reading for the measurement system. The AutoZero will be performed every

time the instrument is turned on. Additionally, you may perform an AutoZero at any time through

the Settings Menu (see Section V.1).
The AutoZero screen will be presented automatically, after the About screen, during a power up.