Rheometer Will Not Return to Zero

Rheometer is not level

• Check with spindle out of the sample.

• Adjust the laboratory stand.

Pivot point or jewel bearing faulty

• Perform an Oscillation Check*

Remove the spindle and turn the motor OFF; select display to % torque mode.

Gently push up on the rheometer coupling.

Turn the coupling until the digital display reads 10-15 on the % display.

Gently let go of the coupling.

Watch the digital display; you should see a “run” of numbers next to the %; the “run” of

numbers should ultimately stop at 0.0 (+/- 0.1).

If the digital display does not return to ZERO, the unit most likely is in need of service.

• Perform calibration check (see Appendix E).

• Contact Brookfield Engineering Laboratories, Inc. or your Brookfield dealer for repair (see

Appendix M).

* This procedure does not apply to instruments with ball bearing suspension (see Section I.3).

Display Reading Will Not Stabilize

Special characteristic of sample fluid. There is no problem with the rheometer.

• Refer to Appendix C.

Check for erratic spindle rotation.

• Verify power supply

• Contact Brookfield Engineering Laboratories, Inc. or your Brookfield dealer for repair.

Bent spindle or spindle coupling.

• Contact Brookfield Engineering Laboratories, Inc. or your Brookfield dealer for repair.

Temperature fluctuation in sample fluid.

• Use temperature bath for control.