beautypg.com

Brookfield DV3T Rheometer User Manual

Page 38

background image

Brookfield Engineering Labs., Inc.

Page 38

Manual No. M13-2100

III.10 Running a Test

A viscosity test is started by pressing the Run button on the Configure Viscosity Test screen. When

Run is pressed, the display will change to the Running Viscosity Test screen (see Figure III-14).

Real Time Graph

Command Keys

Status Bar

Title Bar
Test Name

Measurement Data

Figure III-14

The Running Viscosity Test screen provides information on the current measurement including:

Torque, Viscosity, Shear Stress, Shear Rate, Temperature and Speed.

Torque is the deflection of the Rheometer torque sensor. It is described as a percent (%) and has

a range of 0 – 100%. The DV3T will provide measurement results within the stated accuracy

provided the Torque reading is between 10 and 100%. If the Torque reading falls below 10% the

labels in the data fields will flash to indicate an error condition. Brookfield does not recommend

that data be collected below 10% Torque; however, data collection is not restricted.

TIP: Torque on the DV3T is equivalent to the dial reading from the Brookfield Dial Reading

Viscometer or the % reading from Brookfield Digital Viscometers (DV-E, DV-I,

DV-II, DV-III).

Viscosity is calculated from the measured Torque based on the selected spindle and speed of

rotation. The units of viscosity are defined in the Global Settings section of the Settings Menu

. If the Torque reading falls below 10%, the labels in the data fields will flash to indicate

an error condition.

Shear Stress is calculated from the measured Torque based on the selected spindle. The units of

shear stress are defined in the Global Settings section of the Settings Menu

. If the Torque

reading falls below 10%, the labels in the data fields will flash to indicate an error condition.

TIP: Shear Stress will be displayed as zero for spindles that do not have SRC values.

Shear Rate is calculated from the selected speed based on the selected spindle.

TIP: Shear Rate will be displayed as zero for spindles that do not have SRC values.