beautypg.com

Brookfield DV3T Rheometer User Manual

Page 67

background image

Brookfield Engineering Labs., Inc.

Page 67

Manual No. M13-2100

Regional Settings: The Regional Settings menu displays the current selections for the default

language, number format, date format and day separator. Each setting can

be changed by pressing the blue down arrow and selecting the desired value.

Regional Settings are part of the data file for saved results.

The Language setting established the default language for the DV3T. Individual

users can choose a different language for use with their specific log in (see

Section V.2).

The formats for Number, Time, Day Separator and Date offer several options.

Number Format

9,999,999.00

USA decimal convention

9.999.999,00

Europe comma convention

9 999 999,00

SI/ISO space convention

Time Format

13:59:59

24 Hour Format

1:59:59 PM

12 Hour Format

Day Separator

/

Forward Slash

-

Dash

.

Period

Date Format

Year/Month/Day

ISO Standard

Month/Day/Year

USA Standard

Day/Month/Year

International Standard

Global Alarm: The Global Alarm establishes a single measurement parameter with a range of

acceptance. Whenever the measured value exceeds this range of acceptance,

an audible and visual alarm will be issued. The Global Alarm will apply to any

test that is run. The QC Limit also establishes a measurement parameter with a

range of acceptance; however, the QC Limit is applied only to the test in which

it is defined.

The Global Alarm can be set for any of the four measurement parameters including;

Shear Stress, Temperature, Torque and Viscosity. Once a parameter is selected,

both low and high limits must be entered. These limits are independent of the

spindle and speed selected. When the global alarm is triggered, the DV3T will

provide an audible alarm and a Warning message on the screen (see Figure V-8).

The error message will provide some detail about the alarm condition. The test

and data collection will continue while the Warning message is displayed. The

audible alarm will continue to sound until the operator touches the OK button

within the Warning box.