Spindle List: The selection of Spindle in Configure Viscosity Test can be done by using either a

number pad or a scroll list. (This choice is selected through the navigation button

in the Set Spindle screen.) The contents of the scroll list are defined in the Spindle

List menu. Each spindle can be individually selected or deselected through the use

of a check box. Spindles, shown with both their name and their number, are listed

numerically according to their spindle entry number.

The restricted spindle will not appear on the scroll list of available spindles when

choosing Spindle in Configure Viscosity Test. If the user selects to enter the spindle

by number, then the choice of a restricted spindle will present an Invalid Input

message and prevent the selection of the spindle (see Figure V-10).

List Button

Figure V-10

TIP: To ensure that the correct spindle code is always entered, create a test that

is saved to the DV3T memory.

TIP: The selection of spindle can be restricted by User Log In. Use this restriction

combined with a saved test configuration to ensure that the correct spindle

is always used for critical tests. See Section V.4.2 for details of Users and

Access.

A Special Spindle can be created when the user has developed a new spindle geometry

or when using a standard Brookfield spindle in a non standard measurement container.

The Special Spindle, once created, will be available to users through the Spindle

selection in Configure Viscosity Test.

A Special Spindle is created by selecting the Special Spindle command button in the

Edit Spindle List menu. A Special Spindle requires a code, a name, and an SMC