Brookfield DV3T Rheometer User Manual

Page 33

background image

Brookfield Engineering Labs., Inc.

Page 33

Manual No. M13-2100

Single Point Averaging: Specify an amount of time over which to average measured data.

Collect a single data point when the End Condition is met. This

data point is an averaged value. If the time for averaging is shorter

than the total time for the step, then the average will be performed

for the specified time at the end of the test.

Example 1: The End Condition is Time with a value of 1 minute and 30 seconds, the Single

Point Averaging Duration is 30 seconds, the single data point collected from this

step will be an average of the data measured from 1 minute to 1 minute 30 seconds.

Multi Point: Collect multiple data points based on time. The Data Interval is specified in

Hours:Mins:Secs. If the End Condition is set to Time, then the total number of

points will be calculated and displayed in the Data Collection screen. If the End

Condition is not based on time then it is possible that the step will conclude prior

to a data point being collected. If you want a data point at the conclusion of the

step regardless of the time interval, you can check the check box in the Data

Interval screen.

Example 2: End Condition is Time = 2 minutes, Multi Point Data Interval is 10 seconds. Total

points collected will be 12 with the last data point taken in the last second of the

step.

Example 3: End Condition is set to Viscosity = 200 cP, Multi Point Data Interval is 10 seconds.

During the test the total time required to reach 200 cP is 65 seconds. Total points

collected will be 6 with the last data point taken at 60 seconds, 5 seconds before

the test is finished.

Example 4: End Condition is set to Viscosity = 200 cP, Multi Point Data Interval is 10 seconds.

Check the check box to Also Collect Single Point at Step End. During the test, the

total time required to reach 200 cP is 65 seconds. Total points collected will be 7

with the last data point taken at 65 seconds, 5 seconds after point #6 taken at 60

seconds.

Multi Point Averaging: Specify an amount of time over which to average measured data.

Collect multiple data points based on time until the End Condition is

met. Each data point is an averaged value. If the Averaging Duration

is shorter than the Data Interval for the step, then the average will be

performed for the specified time at the end of the Data Interval. The

total number of points to be collected will be displayed if the End

Condition is set to Time. If the End Condition is not based on time,

then it is possible that the step will conclude prior to a data point

being collected. If you want a data point at the conclusion of the step

regardless of the time interval, you can check the check box in the

Data Interval screen.

Example 5: End Condition is Time = 2 minutes, Multi Point Data Interval is 10 seconds.

Averaging Duration is 5 seconds. Total points collected will be 12 with the last

data point taken in the last second of the step. Each data point will be an average

of the data measured in the last 5 seconds of each Data Interval.

Example 6: End Condition is set to Viscosity = 200 cP, Multi Point Data Interval is 10 seconds.

Averaging Duration is 5 seconds. During the test, the total time required to reach

200 cP is 63 seconds. Total points collected will be 6 with the last data point taken

at 60 seconds, 3 seconds before the test is finished. Each data point will be an

average of the data measured in the last 5 seconds of each Data Interval.

Example 7: End Condition is set to Viscosity = 200 cP, Multi Point Data Interval is 10 seconds.

Averaging Duration is 5 seconds. Check the check box to Also Collect Single Point