3) Total accuracy is the sum of the values n (1) and (2):

At 6 RPM, accuracy is: 15.6 cP + 1.0 cP = +/- 16.6 cP

At 12 RPM, accuracy is: 7.8 cP + 1.0 cP = +/- 9.8 cP

At 30 RPM, accuracy is: 3.1 cP = 1.0 cP = +/- 4.1 cP

4) Therefore, at each speed, the acceptable windows within which the measured viscosity value

must lie is calculated relative to the viscosity value of the standard:

At 6 RPM: 84.9 cP to 118.1 cP

At 12 RPM: 91.7 cP to 111.3 cP

At 30 RPM: 97.4 cP to 105.6 cP

If your measured values fall outside of these windows, contact Brookfield or your authorized

dealer to discuss your results and determine whether your instrument is out of calibration.