Brookfield DV3T Rheometer User Manual

Page 61

background image

Brookfield Engineering Labs., Inc.

Page 61

Manual No. M13-2100

Figure V-3

The creation of a Temperature Offset requires input of the offset value and a name.

The offset value must be in the range of -9.9 to 9.9˚C. The Name can be up to 14

characters long. To delete an existing Temperature Offset, first select the offset from

the list, then press the Delete command key at the bottom of the screen.

The use of a Temperature Offset will be indicated in the Temperature field displayed

in the Configure Viscosity Test screen with a (o) beside the Live indication.

The Temperature menu also allows the user to initiate control over a connected

temperature control device such as the Brookfield Thermosel (HT-106 controller) or

Water Bath (model TC-XX0SD or TC-XX0AP). See Section II.4 for further details.

Establish control over the external temperature controller by selecting Temperature

Controller in the Temperature menu (see Figure V-4).

Figure V-4