Brookfield DV3T Rheometer User Manual
Page 36
Brookfield Engineering Labs., Inc.
Page 36
Manual No. M13-2100
A violation of the QC Limits during the test will be indicated by a flashing yellow box around the
display for the specified parameter. Note that Time does not have a dedicated display. A violation
will also result in an audible beep and a one time warning message (see Figure III-11). Data will
continue to be collected while the warning message is displayed.
Figure III-11
When Viscosity or Torque are the selected QC Limits parameter, QC Limits will be shown during
the test on the Real Time Graph if that parameter is selected for display (see Figure III-12). The
QC Limits are represented by the dashed lines.
TIP: Audible alarms may be turned off in the User Settings .
TIP: An indication of a QC Limit violation is not part of the data set. The user can record a
violation by using the Notes available when viewing test Results (see Section III.11).
Trend Bar
Figure III-12