points.

From the factory 0 counts = 0 force

and 1 mV/V = 5000 lbs force.

7.

Apply the test force for this point

to the loadcell and press ENTER

to calibrate this point. . .

bUSY is displayed while obtaining a

stable value. Pressing ENTER while

bUSY is displayed causes the current

average to be taken immediately instead

of waiting for a stable value. Pressing

any other key while bUSY is displayed

will abort the calibration step being

performed and the display returns to

the previous display.

When a stable value is obtained d n is

displayed. This is a live display of force.

The d is to remind you that this is not

the normal display mode and the n is

the force on the loadcell.

8.

Press ENTER to return to the

point display.

This point is now calibrated.

9.

Repeat steps 6-8 for all the points

you want to calibrate.

Remember, you must press the MENU

key to move to the second calibration

point. A 5000 is the default.

Finish setting up your loadcell by completing the CONFIG. and CLONE parts of

the menu shown in Figure 12. Use Clone to configure a loadcell identically to an

existing loadcell. Just enter the correct loadcell number you want to clone.

After you complete these selections your loadcell will now be remembered as Cell

2. You can now switch back and forth between the two loadcells by connecting

the one you want to the indicator and selecting Cell 1 or Cell 2 using the F1 key.

To exit back to Display mode, press
the
SELECT key and save changes
as needed by pressing
ENTER or
SELECT with SAVE? displayed.

To exit without saving changes, press
Е

Е

Е

Е

Е or MENU until No Save
appears then press
ENTER or
SELECT.