Reset menu / master clear, Figure 20 reset menu – Dillon FI-127 User Manual
Page 47

47
The reset menu shown in Figure 20 appears in two cases.
1.
If you do a Master Clear (powering up the unit with both the MENU and
F2 keys pressed).
2.
If setup, calibration, or data becomes corrupted.
In case 1, you will need to enter the password the same way as explained in the
Service menu. After correctly entering, the reset menu will be displayed.
1. With Reset displayed, press the
ENTER key. . .
The first menu item will be displayed.
See note at left.
2. Press the ENTER key. . .
No is displayed.
3. Choose Yes* to reset to default
values or No to leave the values as
they are. Toggle between the
choices with the Е
Е
Е
Е
Е or MENU key.
When the choice you want is
displayed, press the ENTER key. . .
If you reset the item, the display will
show the next item. When you reset
an item to defaults it disappears from
the menu.
4. Repeat steps 2 and 3 for each item
in the menu. . .
When all items are either accepted or
reset the indicator reboots automati-
cally.
Figure 20
Reset menu
In case 2, the display bypasses the password and goes right to Reset. Repeat steps
1 through 4 above.
Reset Menu / Master Clear
* If the unit is physically sealed, you
must press the internal switch to
select YES.
If the SEAL PHYS. selection is
corrupted, the unit assumes that the
selection is YES.
The reset menu may not contain all
the items shown in Figure 20. If an
item is at default and not corrupted
it will not appear in the menu.
Corrupted items will flash. Items not
at default and not corrupted will
appear but will be solid (not
flashing). These items are not
required to be reset.
You must press the Seal switch if the
unit is physically sealed.
User Master Clear as a last
resort before sending in the
unit for repair.
A software change requires a reset.
You must press the seal switch to
force the reset to occur properly.