Abuse menu figure 7 abuse menu, Figure 8 zero menu, Zero menu – Dillon FI-127 User Manual
Page 20
20
Abuse Menu
Figure 7
Abuse menu
Cell 1 00
This is an example of how a loadcell overload counter appears.
Use the Е
Е
Е
Е
Е or MENU keys to view other loadcells you may
have. The counter shows up to 99 overloads. An overload is
defined as an applied force beyond the threshold configured
under Abused in the loadcell configuration section of the
service menu. The overload counter can only be reset at the
factory.
This menu provides access to zero balance information and reset.
Figure 8
Zero menu
12.3
This is an example of the offset between the current zero
balance and the zero balance value established in calibration.
Displayed offset = Calibration zero - Current zero
When you press the ZERO key the current balance is reset to
calibration zero and resets the peak and 1st peak. The force
display is also unfrozen.
Zero Menu
Be sure to key in the password '127'
before pressing and holding the
ESCAPE key if you want to
change items in the service menu,
otherwise you will only be able to
view them.