beautypg.com

Dillon FI-127 User Manual

Page 23

background image

23

Press the ESCAPE key to exit the test. The outputs remain as
selected until you exit the Test menu. Returning to Display
mode returns the control of the outputs to the cutoff values.

Inputs

This test allows you to check the operation of the onboard and

optional inputs. Press the ENTER key to access the submenu.

Standard

This submenu item lets you check the status of the onboard
inputs. In this example, 1000100, inputs #1 and #5 are active.
The inputs are ordered 1-8 from left to right. A 1 means
activated and a 0 means deactivated.

As you view the inputs, #8 is actually a flag that is dependent on
the states of inputs 6 and 7. Input 6 resets flag #8 to a false 0.
Input 7 sets flag #8 to a true 1. Input 8 does not terminate at a
connector.

Option

This menu item is available only if the option board is installed. It
works the same way as the Standard example above.

Voltages

The submenu under VoltAgES lets you see the power supply

voltages. The voltage readings are updated 2 times per second.

13 volts

This test displays the unregulated weight sensor excitation
power supply voltage. If the voltage drops below 10.5 the display
will show Lo. Volt. The error condition will not clear until
voltage reaches 11.5.

- 5 volts

This test displays the -5 volt excitation voltage.

10 volts

This test displays the unregulated 5 volt logic supply voltage.

24 volts

This test display shows the relay supply voltage. If this voltage
drops too far it may not be possible to activate certain relays.
Nominal level for this power supply voltage is 22.8.

This allows you to test your remote
push button switches (ie. remote
zero, print, etc.)