beautypg.com

Guidance and manufacturer’s declarations, Guidance and, Manufacturer’s declarations – Welch Allyn CardioPerfect Workstation Exercise ECG Module - User Manual User Manual

Page 73: Caution, Electromagnetic emissions

background image

CardioPerfect Workstation

Exercise ECG Module - User Manual

DIR 80012335 Ver. E

73 / 83

13.

Guidance and Manufacturer’s Declarations


Caution

The Welch Allyn CardioPerfect Pro Cardiograph needs special
precautions regarding EMC and needs to be installed and put into
service according to the following EMC information provided.

Portable and mobile RF communications equipment can affect the
Welch Allyn CardioPerfect Pro Cardiograph.



Electromagnetic Emissions

The Welch Allyn CardioPerfect Pro Cardiograph is intended for use in the electromagnetic environment specified

below. The customer or the user of the Welch Allyn CardioPerfect Pro Cardiograph should assure that it is used in

such an environment.

Emissions test

Compliance

Electromagnetic environment

– guidance

RF emissions

CISPR 11

Group 1

The Welch Allyn CardioPerfect Pro Cardiograph uses RF energy
only for its internal function. Therefore, its RF emissions are very
low and are not likely to cause any interference in nearby
electronic equipment.

RF emissions

CISPR 11

Class A

The Welch Allyn CardioPerfect Pro Cardiograph is suitable for
use in all establishments other than domestic, and may be used
in domestic establishments and those directly connected to the
public low-voltage power supply network that supplies buildings
used for domestic purposes, provided the following warning is
heeded:
WARNING: This equipment/system is intended for use by
healthcare professionals only. This equipment/system may
cause radio interference or may disrupt the operation of nearby
equipment. It may be necessary to take mitigation measures,
such as re-orienting or relocating the Welch Allyn CardioPerfect
Pro Cardiograph ro shielding the location.

Harmonic emissions

IEC 61000-3-2

Class A

Voltage fluctuations/
flicker emissions

IEC 61000-3-3

Complies