Multiple portable outlet sockets shall not be placed on the floor. Multiple portable outlet
sockets or extension cord shall not be connected to the system. Do not connect items which
are not part of the system The use of multiple (non-) medical electrical equipment connected
to the same patient may pose a safety hazard due to the summation of leakage currents from
each instrument. Any combination of (non-) medical electrical equipment should be evaluated
by local safety personnel before put into service. Multiple portable outlet sockets use without
an isolation transformer is disapproved unless casual access for additional equipment is
impeded or prevented.

Patient cable and Interface cable

Caution

The patient cable supplied with the CardioPerfect cardiograph is
an integral part of the cardiograph‟s safety features. Using any
other patient cable may compromise defibrillation protection as
well as cardiograph performance. The patient cable should be
routed away from power cords and any other electrical equipment.
Failure to do so can result in AC power line frequency interference
on the ECG trace.

Warning

Do not touch the patient, patient cable, PC interface cable
(Prolink) or cardiograph during defibrillation. Death or injury may
occur from the electrical shock delivered by the defibrillator. It is
recommended to check the patient cable and PC interface cable
(Prolink) for damage prior to the use of the system. If damage
exists do not use the cable, contact your local Welch Allyn Sales
Office or your authorized Welch Allyn dealer or distributor to have
the cable replaced.