beautypg.com

Target workload, O 2.3.13 target workload – Welch Allyn CardioPerfect Workstation Exercise ECG Module - User Manual User Manual

Page 27

background image

CardioPerfect Workstation

Exercise ECG Module - User Manual

DIR 80012335 Ver. E

27 / 83

2.3.13. Target Workload

If an Ergometer was selected as exercise device, the Real-time monitor displays the current
Workload and the current Workload as a percentage of the Target Workload during recording,
in the overview area.

The Target Workload can be adjusted at the start of a new exercise ECG in the New Exercise
ECG window.


The Standard Target Workload shown is calculated with the following formula:

Target Workload = ((2 * Height) - (2.37 * Age) - 73) * 1.01 + 16.6


In case of a female patient multiply the above by 0.8.

Units:
The height is specified in centimeters.
The Age is specified in years.



To set the Target Workload:

Enter a value in the Target Workload field.