beautypg.com

Warranty repair policy – Cooper Instruments & Systems DFI INFINITY B Ultra High Performance Digital Force Indicator User Manual

Page 89

background image

Electrical Safety EN61010-1:2001
Safety requirements for electrical equipment for measurement, control and laboratory.

Double Insulation
Pollution Degree 2
Dielectric withstand Test per 1 min

• Power to Input/Output*:

2500Vac

• Power to Input/Output*:

1500Vac

(Low Voltage dc Power Option**)

• Relays to Power/Inputs/Outputs*:

2500Vac

• Ethernet to Inputs:

1500Vac

• Isolated RS232/RS485 to Inputs:

500Vac

• Isolated Analog to Inputs:

500Vac

Note: * Outputs = Isol. Analog, Isol. RS232/RS485, and Ethernet
Note: **Units configured for external low power dc voltage, 10-32Vdc


Measurement Category I

Category I are measurements performed on circuits not directly connected to the Mains Supply (power).
Maximum measuring voltages to Earth are 120Vdc. This unit should not be used in Measurement Categories II,
III, IV.

Transients Overvoltage Surge (1.2 / 50uS pulse)

Input

Power:

2500V

Input

Power:

1500V

(Low Voltage dc Power Option**)

Ethernet:

1500V

Input/Output

Signals:

500V

Note: **Units configured for external low power dc voltage, 10-32Vdc


EMC EN61326:1997 + and A1:1998 + A2:2001
Immunity and Emissions requirements for electrical equipment for measurement, control and laboratory.

• EMC Emissions Table 4, Class B of EN61326

• EMC Immunity*** Table 1 of EN61326

Note: ***I/O signal and control lines require shielded cables and these cables must be located on conductive
cable trays or in conduits. Furthermore, the length of these cables should not exceed 30 meters

Refer to the EMC and Safety installation considerations (Guidelines) of this
manual for additional information.


WARRANTY REPAIR POLICY

Limited Warranty on Products
Any Cooper Instruments product which, under normal operating conditions, proves defective in material or in
workmanship within one year of the date of shipment by Cooper will be repaired or replaced free of charge provided that a
return material authorization is obtained from Cooper and the defective product is sent, transportation charges prepaid,
with notice of the defect, and it is established that the product has been properly installed, maintained, and operated
within the limits of rated and normal usage. Replacement or repaired product will be shipped F.O.B. from our plant. The
terms of this warranty do not extend to any product or part thereof which, under normal usage, has an inherently shorter
useful life than one year. The replacement warranty detailed here is the buyer’s exclusive remedy, and will satisfy all
obligations of Cooper whether based on contract, negligence, or otherwise. Cooper is not responsible for any incidental
or consequential loss or damage which might result from a failure of any and all other warranties, express or implied,
including implied warranty of merchantability or fitness for particular purpose. Any unauthorized disassembly or attempt to
repair voids this warranty.

Obtaining Service under Warranty

CF 156

84

M2544/N/0505