4 input type: (dc current), 5 input type: (bridge) – Cooper Instruments & Systems DFI INFINITY B Ultra High Performance Digital Force Indicator User Manual
Page 38

VOLt
“VOLt” DC VOLTAGE SUBMENU:
Press ‘▲/MAX’
4) Scroll through the available selection to "VOLt" (flashing).
Press ‘RESET/ENTER’ 5) Display shows "VOLt" again, but not flashing.
Press ‘RESET/ENTER’ 6) Display flashes "UN IPOL" unipolar (default) or previously stored selection of voltage
input type "bI POL" bipolar.
You must choose bi-Polar or uni-Polar before selecting Bridge or Potentiometer
Press ‘▲/MAX’
7) Scroll through the 2 available voltage input types to the selection of your choice.
Press ‘RESET/ENTER’ 8) Display flashes "100 MV"(default) or previous stored input voltage range selection. If the
display matches your choice of Voltage range. Go to Step 10.
Press ‘▲/MAX’
9) Scroll through the available voltage ranges to the selection of your choice (see below
Table for Voltage Input types/Ranges selection).
Press ‘RESET/ENTER’ 10) Display shows "StOREd" stored message momentarily and then advances to the
"RdG.CNF" Reading Configuration Menu.
Display
Voltage types
Voltage Ranges
UN IPOL
100 MV
1 V
10 V
100 V
unipolar
unipolar
unipolar
unipolar
100 milliVolt DC
1 Volt DC
10 Volt DC
100 Volt DC
bI POL
50 MV
500 MV
5 V
50 V
bipolar
bipolar
bipolar
bipolar
± 50 milliVolt DC
± 500 milliVolt DC
± 5 Volt DC
± 50 Volt DC
12.4 Input Type: (DC Current)
ENTER INPUT TYPE MENU:
Press ‘MENU’
1) Display shows "INPUt" Input Menu.
Press ‘RESET/ENTER’ 2) Display flashes "VOLt" (default) or previous stored input type.
Press ‘▲/MAX’
3) Scroll through the available following input type selection list. Display flashes:
"VOLt" (DC Voltage)
"CURRNt" (DC Current)
"bRIdGE" (Bridge)
"POt" (Potentiometer)
"tC" (Thermocouple)
"Rtd" (RTD)
If the displayed input type is "CURRNt", press ‘RESET/ENTER’ to skip to Step 6 (“CURRNt"
stops flashing).
CURRNt
"CURRNt" CURRENT SUBMENU:
Press ‘▲/MAX’
4) Scroll through the available selection to "CURRNt" (flashing).
Press ‘RESET/ENTER’ 5) Display shows "CURRNt" again, but not flashing.
Press ‘RESET/ENTER’ 6) Display flashes "0-20MA" current range (default) or previously stored selection of current
input range (see Table below for available current ranges). If the display matches your
choice of current range. Go to Step 8.
Press ‘▲/MAX’
7) Scroll through the 2 available current input ranges to the selection of your choice.
Press ‘RESET/ENTER’ 8) Display shows "StOREd" stored message momentarily and then advances to the
"RdG.CNF" Reading Configuration Menu.
CURRENT RANGES:
0 - 20 milliampere
4 - 20 milliampere
DISPLAY:
0-20MA
4-20MA
12.5 Input Type: (Bridge)
ENTER INPUT TYPE MENU: