20.2 Ot.SC.OF

Enter OUTPUT SCALE & OFFSET CONFIGURATION Submenu:

READING 1

Press ‘RESET/ENTER

10) Display shows "REAd 1" Reading 1 menu.

Press ‘RESET/ENTER

11) Display flashes 1st digit of previous "REAd 1" value.

Press ‘▲/MAX’ & ‘►/MIN

12) Enter "REAd 1" value. (Example 000000).

Press ‘RESET/ENTER

13) Display advances to "OUt 1" Output 1 menu.

OUTPUT 1

Press ‘RESET/ENTER

14) Display flashes 1st digit of previous "OUt 1" value.

Press ‘▲/MAX’ & ‘►/MIN

15) Enter "OUt 1" value. (Example 0000.00).

Press ‘RESET/ENTER

16) Display advances to "REAd 2" Reading 2 menu.

READING 2

Press ‘RESET/ENTER

17) Display flashes 1st digit of previous "REAd 2" value.

Press ‘▲/MAX’ & ‘►/MIN

18) Enter "REAd 2” value. (Example 999999).

Press ‘RESET/ENTER

19) Display advances to "OUt 2" Output 2 menu.

OUTPUT 2

Press ‘RESET/ENTER

20) Display flashes 1st digit of previous "OUt 2" value.

Press ‘▲/MAX’ & ‘►/MIN

21) Enter "OUt 2" value. (Example 0010.00).

Press ‘RESET/ENTER

22) Display Display shows "StOREd" stored message momentarily and then
advances to the "SP CNF" SetPoint Configuration menu.


The above example is for 0-10 V of the entire range of the process input and analog output. For 0-20mA output
you need to set "MOdE" to "CURRNt" DC-Current output and "OUt 2" to 0020.00.