22.0 SETPOINT CONFIGURATION MENU: (SP.CNF)

Setpoints 1 through 4 can be configured for a very large variety of zone and level signaling.

SP1 and SP2 have balanced configurable hysteresis and are non-latching, suitable for control-level signaling.

SP3 and SP4 are often used as Alarm1 and Alarm2, because they have single-sided hysteresis and can be configured for
latching action.

The levels of these Setpoints are entered during run mode via the front-panel pushbuttons (refer to Setpoint Settings
Menu Configuration Section 8 and many performance options are selected during the configuration mode in the following
sections.

SP.CNF

ENTER SETPOINT CONFIGURATION MENU:

Press ‘MENU

1) Six times, display shows "SP.CNF" Setpoint Configuration Menu.

Press ‘RESET/ENTER

2) Display flashes "dISAbL" (default) or previous selection.

Press ‘▲/MAX

3) Select "ENAbLE" to configure Setpoint 1 and Setpoint 2.

SP1.CNF

Press ‘RESET/ENTER

4) Display shows "StOREd" stored message momentarily and then advances to "SP
1.CNF
" Setpoint 1 configuration submenu. If necessary, press ‘MENU’ to select the other
available following submenu: "SP 2.CNF" Setpoint 2 Configuration and "SP db" Setpoint
Deadband Configuration Submenu. See Steps 12 through 16 for details.

Press ‘RESET/ENTER

5) Display shows "ACt IVE" Setpoint Active State menu.

Press ‘RESET/ENTER

6) Display flashes "AbOVE" Above Setpoint active state option or previous setting.

Press ‘▲/MAX

7) Scroll to select 2 available choices: "AbOVE" or "bELOU" (below) Setpoint Active
State.

Press ‘RESET/ENTER

8) Display shows "StOREd" stored message momentarily and then advances to
"N.StAtE" Normal State (Open-collector or Relay Output active state) menu.

Press ‘RESET/ENTER

9) Display flashes "N.OPEN" Normal Open (Output collector or Relay opened) state or
previous setting "N.CLOSE" Normal Close (Output Relay closed) state.

Press ‘▲/MAX

10) Select your choice of Output Relay state setting.

Press ‘RESET/ENTER

11) Display shows "StOREd" stored message momentarily and then advances to "SP
2.CNF
" Setpoint 2 Configuration Menu.

SP2.CNF

Press ‘RESET/ENTER’ &
‘▲/MAX

12) "SP 2.CNF" Setpoint 2 Configuration submenu items are identical to "SP 1.CNF",
proceed with your choices of setting then display advances to "SP db" Setpoint Deadband
Menu.

Press ‘RESET/ENTER

13) Display shows "StOREd" stored message momentarily and then advances to "SP db"
Setpoint Deadband Configuration Submenu. Setpoint Deadband is used to select the
amount of hysteresis for the setpoints programmable from "0000" to "9999").

SP db

Press ‘RESET/ENTER

14) Display shows some 4-digit number (0020 as default) of previous setting with the first
digit flashing.

Press ‘▲/MAX’ & ‘►/MIN’ 15) To modify the value of Setpoint Deadband.
Press ‘RESET/ENTER

16) Display shows "StOREd" stored message momentarily and then advances to "AL
CNF
" Alarm Configuration Menu.


CF 156

53