Figure 5-8. 4 Relay Output Option Board


The table below shows which jumpers are assigned to each relay. Defaults have asterisks.

Table 5-3 4 Relay Board Jumper Selection

S1 S2

FUNCTION

A, C*

A, C*

Assigns SP1 to Relay 1 (P6)
Assigns SP2 to Relay 2 (P7)
Assigns SP3 to Relay 3 (P18A)
Assigns SP4 to Relay 4 (P18B)

B, D

A, C

Assigns SP1 to Relay 3 (P18A)
Assigns SP2 to Relay 2 (P7)
Assigns SP3 to Relay 1 ( P6)
Assigns SP4 to Relay 4 (P18B)

B, D

B, D

Assigns SP1 to Relay 3 (P18A)
Assigns SP2 to Relay 4 (P18B)
Assigns SP3 to Relay 1 ( P6)
Assigns SP4 to Relay 2 ( P7)

A, C

B, D

Assigns SP1 to Relay 1 (P6)
Assigns SP2 to Relay 4 (P18B)
Assigns SP3 to Relay 3 (P18A)
Assigns SP4 to Relay 2 (P7)

Figure 5-9. Analog Output Option Board

5.2.4 Panel-Mount Assembly

The meter can be mounted in a panel so that the front of the meter is flush with the vertical panel surface. Panel mounting
can be seen as simply “sandwiching” the panel between the inner case and the outer sleeve in the last phases of

CF 156

14

M2544/N/0505