Al 2.cnf enter alarm 2 configuration menu, 3 al db enter alarm deadband menu, 4 num.dly enter number of readings/display menu – Cooper Instruments & Systems DFI INFINITY B Ultra High Performance Digital Force Indicator User Manual
Page 63: 5 rst al enter alarm reset control menu

Press ‘RESET/ENTER’ 15) Display flashes "PROC." Process (No Deviation) mode or previous selection.
Press ‘▲/MAX’
16) Scroll 4 available following options for your choice of Alarm Mode: "PROC." (Process
Deviation), "HI dEV" (High Deviation), "LO dEV" (Low Deviation) and "bNd.dEV" (Band
Deviation).
24.2. AL 2.CNF
Enter ALARM 2 CONFIGURATION Menu:
Press ‘RESET/ENTER’ 17) Display shows "StOREd" stored message momentarily and then advance to "AL
2.CNF" Alarm2 configuration submenu. If necessary, Press ‘MENU’ to skip and enter the
other available following submenu: "AL db" Alarm Deadband Configuration, "NUM.dLy"
Number of display/reading configuration, "RSt AL" Alarm Reset Submenu.
Press ‘RESET/ENTER’ 18) To enter "AL 2.CNF" Alarm 2 configuration which all menu items are identical with
"AL 1.CNF" Alarm1 configuration menu. Use ‘RESET/ENTER’ and ‘MAX’ buttons to
setup Alarm 2 Functions or Press ‘MENU’ to skip it and enter "AL db" Alarm Deadband
setup menu.
24.3 AL db
Enter ALARM DEADBAND Menu:
This selected deadband (hysteresis) for Alarm 1 ("SP3") and Alarm 2 ("SP4") is placed on the INACTIVE side of the
selected levels. This results in immediate action (if zero (0) delay is selected in "NUM.dLy") When an alarm limit is
exceeded, but defers recovery when the input returns to pre-alarm levels.
Press ‘MENU’
19) Display shows "AL db" Alarm Deadband setup menu.
Press ‘RESET/ENTER’
20) Display shows some 4-digit number (0020 as default) of previous settings with the
first digit flashing.
Press ‘▲/MAX’ & ‘►/MIN’ 21) To modify the value of Alarm Deadband for your desired setting.
Press ‘RESET/ENTER’
22) Display shows "StOREd" stored message momentarily and then advances to
"NUM.dLy" Number of input readings required triggering Alarm 1 (SP3) and Alarm 2
(SP4) action.
24.4 NUM.dLy
Enter NUMBER OF READINGS/DISPLAY Menu:
Press ‘RESET/ENTER’ 23) Display shows two 2-digit numbers (03 03 as default) with the second right side of
display) one flashing.
Press ‘▲/MAX’
24) To modify the delay value or number of input readings before Alarm 2 response.
Press ‘►/MIN’
25) To select Alarm 1’s delay value setting, display flashing the first 2-digit number (left
side of the display). Pressing ‘MIN’ button allow users to select Alarm 1 or Alarm 2’s
delay values.
Press ‘▲/MAX’
26) To modify the delay value or number of input readings before Alarm 1 response.
Press ‘RESET/ENTER’ 27) Display shows "StOREd" stored message momentarily and then advances to "RSt
AL" Alarm Reset submenu.
24.5 RSt AL
Enter ALARM RESET CONTROL Menu:
Press ‘RESET/ENTER’
28) Display flashes "dISAbL" (disable) or previous setting.
Press ‘▲/MAX’
29) To enable "ENAbLE" or disable "dISAbL" both Alarm 1 and Alarm 2 (Setpoints 3
and 4) action and LEDs.
Press ‘RESET/ENTER’
30) Display shows "StOREd" stored message momentarily and then advances to
"COMM" Communication parameter setting menu.