beautypg.com

Notes, warnings and cautions, 0 introduction, 1 description – Cooper Instruments & Systems DFI INFINITY B Ultra High Performance Digital Force Indicator User Manual

Page 6: 2 features, 3 available options and accessories

background image

NOTES, WARNINGS AND CAUTIONS

Information that is especially important to note is identified by these labels:

• NOTE

• WARNING

• CAUTION

• IMPORTANT


NOTE: provides you with information that is important to successfully set up and use the Programmable Digital meter.
CAUTION or WARNING: tells you about the risk of electric shock.
CAUTION, WARNING or IMPORTANT: tells you of circumstances or practices that can affect the meter’s functionality and
must refer to accompanying documents.

1.0 INTRODUCTION

1.1 Description

This versatile ultra high-performance meter handles a broad spectrum of dc voltage and current ranges, nine
thermocouple types, multiple RTD's, and signals from strain gauge transducers such as load cells and pressure
transducers, as well as potentiometric inputs. It also features ten point linearization of input signals, programmable by the
user for custom applications.

Built-in excitation to power virtually any sensor or transmitter, and four isolated open collector outputs for control or alarms
are standard. The big, bright, 6-digit patented LED display can be programmed to change color between Red, Amber, and
Green at any setpoint. Output options include: Isolated programmable analog voltage or current and four relays.

With the Serial Communications option, the user can select from a push-button menu between RS-232, RS-485, and
either a straightforward ASCII protocol or MODBUS. With our award-winning Ethernet/Internet option, this unit connects
directly to an Ethernet network and transmits data in standard TCP/IP protocol. It is possible to monitor and control a
process through a web browser from anywhere on the Internet.

1.2 Features

• Accuracy: up to ±0.005% of reading

• 6-Digit Color Changing LED Display

• Up to 166 Readings Per Second

• 10 Point Linearization

• 4 Isolated Open Collector Output

• Isolated Analog Output, Optional

• Four Relays, Optional

• Ethernet or RS-232/RS-485, Optional

• Low Voltage Power, Optional

• In-Line Calibration

• NEMA 4 Front Bezel

1.3 Available Options and Accessories

The following options are available. Optional boards are either installed at the time of purchase or available as separate
items and installed by the user after purchase.

-T/C

-4R

-A

-C24

-EI

-DC

Thermocouple Temperature Compensation Connector Kit
Four Form-C SPDT Relay Output Board
Isolated 14 bit Analog Output Board
Serial Communications RS-232* + RS-485** + MODBUS
Ethernet/Internet
Low Voltage Power Option

CF 156

1

M2544/N/0505