Cooper Instruments & Systems DFI INFINITY B Ultra High Performance Digital Force Indicator User Manual
Page 69

button.
• For Universal meters, there is a list of the transmitted character (Hex number value)
corresponding for each of the sixteen (16) possible Setpoint/Alarm (SP1, SP2, AL1/SP3
and AL2/SP4) states. Refer to STATUS CHARACTER FORMATS/Alarm Status
Characters Chapter of Serial Communications Option Manual for detail description and
listing Table.
Press ‘RESET/ENTER’ 34) Display flashes "NO" No (Alarm Status Character excluded from Data String as
default) option or previous selection.
Press ‘▲/MAX’
35) Scroll or select either "NO" to exclude or "yES" to include Alarm Status Character in
Data String Format for your choice of setting.
P.V.StAt
Press ‘RESET/ENTER’ 36) Display shows "StOREd" stored message momentarily and then advances to the
"P.V.StAt" Peak & Valley Character Setting Submenu.
• The meter keeps track of Peak and Valley changes at the completion of each
measurement, so the occurrence of new values can be signaled. However,
transmissions of Peak/Valley status may be commanded less often: therefore, the meter
also monitors if Peak or Valley values have changed since the last status transmission.
• These Characters only applies for Universal meters. Refer to STATUS CHARACTER
FORMATS/Peak & Valley (Hi/Lo) Status Characters Chapter of Serial Communications
option Manual for detail description and listing Table.
Press ‘RESET/ENTER’ 37) Display flashes "NO" No (the Character Excluded as default) option or previous
selection.
Press ‘▲/MAX’
38) Scroll and select either "NO" or "yES" to exclude or include the character from Data
String Format.
SNd.dAt
Press ‘RESET/ENTER’ 39) Display shows "StOREd" stored message momentarily and then advances to the
"SNd.dAt" Current Reading Value Transmission submenu that allows user to select
whether or not the data string transmitted is included with current reading.
Press ‘RESET/ENTER’ 40) Display flashes, "yES" as included the current reading value in data string as default
or previous selection.
Press ‘▲/MAX’
41) Scroll and select either "yES" to include or "NO" to exclude current reading in data
string transmitted.
FL.tRNS
Press ‘RESET/ENTER’ 42) Display shows "StOREd" stored message momentarily and then advances to the
"FL.tRNS” Filtered Data Transmission Switch Submenu that allows user to select whether
or not filtered data to be transmitted as part of Data String.
Press ‘RESET/ENTER’ 43) Display flashes "NO" No (not transmitted as default) option or previous selection.
Press ‘▲/MAX’
44) Scroll and select "NO" to exclude or "yES" to include filtered data transmission for
choice of setting.
PK.tRNS
Press ‘RESET/ENTER’ 45) Display shows "StOREd" stored message momentarily and then advances to the
"Pk.tRNS" Peak Value Transmission submenu that allows user to select whether Peak
Value to be included or excluded in the Date String.
Press ‘RESET/ENTER’ 46) Display flashes "NO" No (excluded as default) option or previous selection.
Press ‘▲/MAX’
47) Scroll and select "NO" to exclude or "yES" to include Peak Value in the Data String for
choice of setting.
VL.tRNS
Press ‘RESET/ENTER’ 48) Display shows "StOREd" stored message momentarily and then advances to the
"VL.tRNS" Valley Value Transmission submenu that allows user to select whether Valley
Value to be included or excluded in the Date String.
Press ‘RESET/ENTER’ 49) Display flashes "NO" No (excluded as default) option or previous selection.
Press ‘▲/MAX’
50) Scroll and select "NO" to exclude or "yES" to include Valley Value in the Data String
for choice of setting.
CF 156
64
M2544/N/0505