0 jumper settings, 1 thermocouple input, 2 rtd input – Cooper Instruments & Systems DFI INFINITY B Ultra High Performance Digital Force Indicator User Manual
Page 21: 1 for rtds, 2 for rtd 10 ohm copper

11. Set P1, the AC power connector, aside and connect or reconnect all other connectors to the back of the meter using
Figure 4-3 in Section 4.2 as guides.
Connect P1 last.
The P1 connector is
“keyed”; it is shaped in such a way that it fits only the J1 male pins.
12. Replace the rear protective cover and secure it with the cover mounting screw.
Your meter is now ready for operation and you can turn-on the power.
WARNING: The unit becomes active as soon as it is connected to a power source. It has no On or Off switch.
The unit at first momentarily shows the software (or Firmware) version number, during this time, user can
switch to older Configuration Menu, “MENU 2” (the use of MENU 2 is not covered in this manual, refer to
“Menu 2 Manual’ for proper instructions), followed by “RESET 2”, and then proceeds to the Run Mode.
6.0 JUMPER SETTINGS
This section is for the configuration and setup of your jumper positions for readrate, unipolar or bipolar signal input, sensor
input signal jumpers, sensor excitation jumpers, pushbutton lockouts and lockout of lockout configuration menus.
Figure 6-1. S1 and S2 Jumper Locations on Signal Input Board
6.1 THERMOCOUPLE Input:
S1_G/J; S2_A/B/K, jumpers are closed as shown:
6.2 RTD Input:
6.2.1 For RTDs:
S1_H; S2_A/B/F/J, jumpers are closed as shown:
6.2.2 For RTD 10 ohm copper:
S1_H; S2_A/B/F/J/L, jumpers are closed as shown:
CF 156
16
M2544/N/0505