Input Signal Range
Scale Input 1 Value
Scale Input 2 Value
0-50mV
000000
050000
0-100mV
000000
100000
0-1V
000000
100000
0-5V
000000
050000
1-5V
010000
050000
1-6V
010000
060000
Input Signal Range
Scale Input 1 Value
Scale Input 2 Value
0-10V
000000
100000
0-50V
000000
050000
0-100V
000000
100000
±5V
-50000
050000
±50V
-50000
050000
Table 16-2. Factory Calculated Scale Factors Table
For the purpose of this procedure, we will use an input of 0-50 mVdc and scale the meter to display 0 to 2500.0. The
calculated scale factors can now be used to scale your meter to display in engineering units using Input Scale and Offset
("IN.SC.OF") as follows:
Record your "INPUT1" and “READ 1" numbers, and your "INPUT2" and "READ 2" numbers, record them below and
proceed as follows:
Example
INPUT1 = 000000
READ 1 = 000000
INPUT2 = 050000
READ 2 = 025000
For more detail of Scale & Offset
’s examples, formulas and Calculation Table, Refer to Section 18 of MULTI-POINT
SCALE & OFFSET MENU CONFIGURATION.