2 input type: (rtd ), 3 input type: (dc voltage/process value) – Cooper Instruments & Systems DFI INFINITY B Ultra High Performance Digital Force Indicator User Manual
Page 37

below for types). If the display matches your choice of Thermocouple. Go to Step 8.
Press ‘▲/MAX’
7) Scroll through the available thermocouple types to the selection of your choice.
Press ‘RESET/ENTER’ 8) Display shows "StOREd" stored message momentarily and then advances to the
"RdG.CNF" Reading Configuration Menu.
Thermocouple
Types:
J K T E N DIN
J R S B
Display: J tC k tC t tC E tC N tC dNJ tC R tC S tC b tC
12.2 Input Type: (RTD )
ENTER INPUT TYPE MENU:
Press ‘MENU’
1) Display shows "INPUt" Input Menu.
Press ‘RESET/ENTER’ 2) Display flashes "VOLt" (default) or previous stored input type.
Press ‘▲/MAX’
3) Scroll through the available following input type selection list. Display flashes:
"VOLt" (DC Voltage)
"CURRNt" (DC Current)
"bRIdGE" (Bridge)
"POt" (Potentiometer)
"tC" (Thermocouple)
"Rtd" (RTD)
If the displayed input type is "Rtd", press ‘RESET/ENTER’ to skip to Step 6 ("Rtd" stops
flashing).
Rtd "Rtd"
RTD
SUBMENU:
Press ‘▲/MAX’
4) Scroll through the available selection to "Rtd" (flashing).
Press ‘RESET/ENTER’ 5) Display shows "StOREd" stored message momentarily and then "Rtd" (not flashing).
Press ‘RESET/ENTER’ 6) Display flashes previous RTD type selection. ie. "4Pt 385" default (see Table below for
types). If the display matches your choice of RTD type. Go to Step 8.
Press ‘▲/MAX’
7) Scroll through the available RTD types to the selection of our choice.
Press ‘RESET/ENTER’ 8) Display shows "StOREd" stored message momentarily and then advances to the
"RdG.CNF" Reading Configuration Menu.
Display RTD
types
2Pt 392
3Pt 392
4Pt 392
2 wire, 100 Ohms, NIST RTD, alpha=0.00392
3 wire, 100 Ohms, NIST RTD, alpha=0.00392
4 wire, 100 Ohms, NIST RTD, alpha=0.00392
2Pt 385
3Pt 385
4Pt 385
2 wire, 100 Ohms, NIST RTD, alpha=0.00385
3 wire, 100 Ohms, NIST RTD, alpha=0.00385
4 wire, 100 Ohms, NIST RTD, alpha=0.00385
3.L INR
4.L INR
3 wire, linear RTD
4 wire, linear RTD
12.3 Input Type: (DC Voltage/Process Value)
ENTER INPUT TYPE MENU:
Press ‘MENU’
1) Display shows "INPUt" Input Menu.
Press ‘RESET/ENTER’ 2) Display flashes "VOLt" (default) or previous stored input type.
Press ‘▲/MAX’
3) Scroll through the available following input type selection list. Display flashes:
"VOLt" (DC Voltage)
"CURRNt" (DC Current)
"bRIdGE" (Bridge)
"POt" (Potentiometer)
"tC" (Thermocouple)
"Rtd" (RTD)
If the displayed input type is "VOLt", press ‘RESET/ENTER’ to skip to Step 6 ("VOLt" stops
flashing).
CF 156
32