4 modbus enter modbus protocol submenu, 5 addres enter address selection submenu, 6 sercnt enter serial count submenu – Cooper Instruments & Systems DFI INFINITY B Ultra High Performance Digital Force Indicator User Manual
Page 70

SEPARA
Press ‘RESET/ENTER’ 51) Display shows "StOREd" stored message momentarily and then advances to the
"SEPARA" Separator Selection submenu that determines whether different data types
included in Data String and sent from the device will be separated by spaces or by
Carriage Returns.
Press ‘RESET/ENTER’ 52) Display flashes "SPACE" Space separator as default or previous selection.
Press ‘▲/MAX’
53) Scroll and select "SPACE" or "CR"
Data String Format.
UNIt
Press ‘RESET/ENTER’ 54) Display shows "StOREd" stored message momentarily and then advances to the
"UNIt" Unit of Measure submenu that allows user to include user-Defined unit of
measurement in Data String Format (For detail and description of how to create your own
units of measurement, refer to Serial Communication Operation Manual).
Press ‘RESET/ENTER’ 55) Display flashes "NO" No (excluded as default) option or previous selection.
Press ‘▲/MAX’
56) Scroll and select "NO" to exclude or "yES" to include your own defined unit of
measure in the Data String for your choice of Data Format.
26.4 MOdbUS
Enter MODBUS PROTOCOL Submenu
Press ‘RESET/ENTER’ 57) Display shows "StOREd" stored message momentarily and then advances to the
"MOdbUS" Modbus Protocol submenu.
Press ‘RESET/ENTER’ 58) Display flashes "NO" No option as default as Modbus Protocol disabled or previous
selection.
Press ‘▲/MAX’
59) Scroll and select "NO" to disable or "yES" to enable Modbus protocol.
• If Modbus Protocol was selected, the following Communications Parameters must be set
as: No Parity, 8-bit Data Bit, 1-Stop Bit. Do not attempt to change these parameters.
26.5 AddRES
Enter ADDRESS SELECTION Submenu
Press ‘RESET/ENTER’
60) Display shows "StOREd" stored message momentarily and then advances to the
"AddRES” Address selection submenu.
• This menu is applicable to the RS-485 Option only to designate different address for
each device or controller in the Multi-point Serial Communication network that allows
hosting device to recognize and communicate properly as labeling or naming
purposes.
Press ‘RESET/ENTER’
61) Display flashes "001" Address Value 1 as default.
Press ‘▲/MAX’ & ‘►/MIN’
62) To modify and select Address Value for your choice of setting. This controller
supports value within 000
– 199.
26.6 SERCNt
Enter SERIAL COUNT Submenu
Press ‘RESET/ENTER’
63) Display shows "StOREd" stored message momentarily and then advances to the
"SERCNt" Serial Count submenu.
• Serial Count menu is used to program the number of readings the meter just take
(programmable from "00001" to "59999") between transmissions of data via the Serial
Communication Option Board.
Caution: For proper communication, do not set Serial Count Value at "00000".
Press ‘RESET/ENTER’
64) Display flashes "00001" as default Serial Count value.
Press ‘▲/MAX’ & ‘►/MIN’
65) To modify and select the desired value of Serial Count (number of readings)
between data transmissions.
Press ‘RESET/ENTER’
66) Display shows "StOREd" stored message momentarily and then advances to the
"COLOR" LED Color Select submenu.
CF 156
65
M2544/N/0505