SEPARA

Press ‘RESET/ENTER 51) Display shows "StOREd" stored message momentarily and then advances to the

"SEPARA" Separator Selection submenu that determines whether different data types
included in Data String and sent from the device will be separated by spaces or by
Carriage Returns.

Press ‘RESET/ENTER 52) Display flashes "SPACE" Space separator as default or previous selection.
Press ‘▲/MAX

53) Scroll and select "SPACE" or "CR" Carriage Return Option for your choice of
Data String Format.

UNIt

Press ‘RESET/ENTER 54) Display shows "StOREd" stored message momentarily and then advances to the

"UNIt" Unit of Measure submenu that allows user to include user-Defined unit of
measurement in Data String Format (For detail and description of how to create your own
units of measurement, refer to Serial Communication Operation Manual).

Press ‘RESET/ENTER 55) Display flashes "NO" No (excluded as default) option or previous selection.
Press ‘▲/MAX

56) Scroll and select "NO" to exclude or "yES" to include your own defined unit of
measure in the Data String for your choice of Data Format.

26.4 MOdbUS

Enter MODBUS PROTOCOL Submenu

Press ‘RESET/ENTER’ 57) Display shows "StOREd" stored message momentarily and then advances to the

"MOdbUS" Modbus Protocol submenu.

Press ‘RESET/ENTER’ 58) Display flashes "NO" No option as default as Modbus Protocol disabled or previous

selection.

Press ‘▲/MAX

59) Scroll and select "NO" to disable or "yES" to enable Modbus protocol.

• If Modbus Protocol was selected, the following Communications Parameters must be set

as: No Parity, 8-bit Data Bit, 1-Stop Bit. Do not attempt to change these parameters.

26.5 AddRES

Enter ADDRESS SELECTION Submenu

Press ‘RESET/ENTER

60) Display shows "StOREd" stored message momentarily and then advances to the
"AddRES” Address selection submenu.

• This menu is applicable to the RS-485 Option only to designate different address for

each device or controller in the Multi-point Serial Communication network that allows
hosting device to recognize and communicate properly as labeling or naming
purposes.

Press ‘RESET/ENTER

61) Display flashes "001" Address Value 1 as default.

Press ‘▲/MAX’ & ‘►/MIN

62) To modify and select Address Value for your choice of setting. This controller
supports value within 000

– 199.

26.6 SERCNt

Enter SERIAL COUNT Submenu

Press ‘RESET/ENTER

63) Display shows "StOREd" stored message momentarily and then advances to the
"SERCNt" Serial Count submenu.

• Serial Count menu is used to program the number of readings the meter just take

(programmable from "00001" to "59999") between transmissions of data via the Serial
Communication Option Board.

Caution: For proper communication, do not set Serial Count Value at "00000".

Press ‘RESET/ENTER

64) Display flashes "00001" as default Serial Count value.

Press ‘▲/MAX’ & ‘►/MIN

65) To modify and select the desired value of Serial Count (number of readings)
between data transmissions.

Press ‘RESET/ENTER

66) Display shows "StOREd" stored message momentarily and then advances to the
"COLOR" LED Color Select submenu.


CF 156

65