beautypg.com

Single, External trigger, Gate delay – Bird Technologies SH-362S-Manual User Manual

Page 88: Trigger arm trigger, External - low level / high level

background image

66

Single
Performs a single sweep. Before a single sweep is triggered, the SignalHawk will
display the previous sweep. To begin a sweep, press the Manual Trigger soft
key. After the sweep finishes, it will display the results and stop sweeping.

Trigger -

Sets the type of trigger that initiates a sweep. There are four types of

trigger to choose from:

Internal Continuous - Sweeps continuously and is controlled by the Sig-
nalHawk.

Internal Single - Sweeps once when triggered. Controlled by the Signal-
Hawk through the Arm Trigger soft key.

External Continuous - See “External Trigger” below.

External Single - See “External Trigger” below.

Note:

Useful for signals that are time varying, such as ones that

change slow enough to trigger manually as well as carrier/interfer-
ence ratios. See “Carrier-to-Interference Ratio” on page 94.

Arm Trigger -

Manually initiates a sweep.

External Trigger

External - Low Level / High Level /
Rise Edge / Fall Edge / Either Edge

Low Level - Triggers if input = 0 ± 0.5 V (TTL “Low”).

High Level - Triggers if input = 4.2 ± 0.8 V (TTL “High”).

Rise Edge - Triggers if the input goes from “Low” to “High”.

Fall Edge - Triggers if the input goes from “High” to “Low”.

Either Edge - Triggers if the input goes from either “High” to “Low” OR
“Low” to “High”.

Gate Delay

Length of wait after the trigger signal and before beginning a sweep. The range
is 100 μs to 1 s.

This manual is related to the following products: