beautypg.com

Vector network analyzer quick start – Bird Technologies SH-362S-Manual User Manual

Page 40

background image

18

Vector Network Analyzer Quick Start

1.

Measure the output power of the system being tested, or signal power, at
the system’s test port using a power meter, service monitor, or equivalent.

Note:

Ensure the output power is less than +22 dBm (160 mW).

2.

Power up the SignalHawk.

3.

In the Start Menu, use the arrow keys to highlight the desired measure-
ment and press Enter.

4.

In the Freq & Span Menu, set the frequency range (See “Freq & Span
Menu” on page 58).

5.

Do one of the following:

For Match or Cable Loss Measurement:
a. Select either the Match or Cable Loss measurement.

b. Press Calibrate

c. Select Open, Short, or Load (soft keys) on RF Out corre-

sponding with a calibrated open, short, or load connected

to the RF Out port.

d. Connect the cable and antenna to the RF Out test port.

Note:

Maximum input is +22 dBm (160 mW).

CAUTION

Vector Network Analyzer has a +22 dBm (160 mW) max. RF

input. Exceeding the maximum input will damage the

SignalHawk. If unsure of power levels, measure the test

connection with a power sensor before using the

SignalHawk.

CAUTION

Do not turn on a DC bias if the thru cable is attached from

the VNA In to the VNA Out. This will damage the VNA input

and destroy the unit.

CAUTION

Center conductor cabling should be discharged prior to

connecting. See “Discharging Static Electricity” on page 131.

Any large electrostatic discharge or high power RF applied

directly to the RF input or output will cause internal damage

and void the warranty.

Power to any adjacent broadcast antennas should be shut

down as well. If adjacent antenna broadcasts cannot be

terminated, then adding sufficient external attenuation to the

input of the instrument prior to connecting is recommended.

Failure to do so could result in irreparable damage to the

SignalHawk.

This manual is related to the following products: