Units, Vswr, Return loss – Bird Technologies SH-362S-Manual User Manual

Page 52: Match, Reflection, Pwr out, Sweep, Cont/single, Data points, Vswr return loss rho match % reflection

background image

30

Units

VSWR

Return Loss

rho

Match %

Reflection %

Pwr Out

Controls the power level of the signal that is sent out to the RF port. The Signal-
Hawk then measures the power that comes back. The default setting is 0 dBm. Can
be set from -20 to +10 in 1 dB steps. This should be set as high as possible without
overpowering the system.

Note:

For most measurements, the output should be set to 0dBm or

lower.

Note:

For systems with good match (low reflected power) set the

output higher to improve the signal to noise ratio.

Sweep

Sets the parameters of the sweep for the SignalHawk.

Cont/Single
Toggles the sweep setting from a single sweep to a continuous cycle of sweep-
ing and recording.

Note:

Selecting this will enable a manual trigger soft key that starts

the sweep when pressed.

Data Points
Controls the sweep speed and level of detail. The range can go from 12 to 11265
points per sweep.

Note:

Increasing the number of data points increases the measur-

able distance in Distance-to-Fault measurements and increases the
detail in Measure Match mode measurements. Use a smaller number
of points when frequency resolution is not critical & a faster sweep
time is desired. Use more points for a more accurate display of finer
frequency divisions.

1

Sqrt Pr Pf

+

 1 Sqrt Pr Pf

dB

10

Log Pr Pf

=

Sqrt Pr Pf

100

1

Pr Pf

100

1

Pr Pf